Universal 3-Wire Input Adapter
Single Output Switching Power Supply
Veridiam Tissue Arrayer instrumentation comes with either a 120-volt or a 240-volt Power Supply.
Output Specifications
| Model |
DC Output Voltage |
Min. Load |
Max. Load |
Ripple (1) P-P (max.) |
Regulation L-L |
| 120-volt |
+ 12V |
0A |
2.5A |
120mV |
±5% |
| 240-volt |
+ 24V |
0A |
1.25A |
240mV |
±5% |
(1) Measured with by-pass capacitors 0.1uf/10uf at output connector terminal and oscilloscope set at 20 Mhz. Phihong is not responsible for any error and change. Please consult the factory and visit www.phihong.com for the most up-to-date specifications.
| Features |
| • Low Cost |
• Class B EMI |
• Non-vented/Spill-proof Case |
• Low Profile Design |
| Safety Approvals |
| • cUL/UL |
• CE |
• C-Tick |
• TUV |
• SAA |
• Meeting Cal. Prop. 65 |
| Mechanical Characteristics |
• Length: 4.45 in. (113 mm) |
• Width: 2.4 in. (61 mm) |
• Height: 1.34 in. (34 mm) |
• Weight: 8.5 oz. (242 g) |
| Input Characteristics |
• AC Input Voltage Rating: 120 to 240VAC |
• AC Input Voltage Range: 90 to 264VAC |
• AC Input Frequency: 47 to 63Hz |
• Input Current: 1A (rms) Max. @ 115VAC 0.5A (rms) Max. @ 230VAC |
• Leakage Current: 3.5mA Max. |
• Inrush Current (cold): 30A for 115VAC at Max. load 60A for 230VAC at Max. load (Cold start @ ambient 25°C) |
| Output Characteristics |
• Efficiency: 70% typical |
• Hold-up Time: 12mS Min. @ 120VAC and Max. load |
• Over-voltage Protection: >130% over-voltage detection, adapter will be shut down |
• Over-current Protection: Output equipped with short circuit protection - auto
restart |
• Short Circuit Protection: Output can be shorted without damage |
|
| Environmental Characteristics |
• Tempeature: Operation 0 to +45°C Non-operation -30 to +85°C |
• Humidity: Operation 8 to 90% |
• Emissions: FCC Class B EN55022 Class B VCCI Class B |
• Immunity: 61000-4-2 Level 4 ESD 61000-4-4 Level 2 FET/Burst 61000-4-5 (L-N) Level 2 Surge (L-G) Level 3 |
• Dielectric Withstand (Hi-pot) Test: Primary to Secondary 1500VAC for 1 minute, 10mA |
DC Output Connector: 2.1x5.5mm Center Positive Standard |